Module 04 · Measurement pitfalls
Measurement pitfalls
Seven short stations. Each one lets you break a measurement with your own hands, see what the monitor shows, and learn the check that would have caught it.
Zero & level
Station 1 of 7
Move the transducer up and down the pole and watch what the monitor reports for a right atrial pressure that is truly 4 mmHg.
belowat axisabove
Displayed mean
4.0mmHg
True mean
4.0mmHg
Error
0.0mmHg
0 cm × 0.74
displayed = true − (height above axis in cm × 0.74)
1 cm H₂O ≈ 0.74 mmHg (1 mmHg = 1.36 cm H₂O) · 0 cm → 0.0 mmHg
Quick check Self-test
The transducer is 10 cm below the phlebostatic axis. The monitor reads an RA mean of 11 mmHg. What is the true RAP?
Why this matters
- Zero the transducer to atmosphere, then level it at the phlebostatic axis: the 4th intercostal space in the mid-axillary line, about 5 cm below the sternal angle in the supine patient. Kern 2022Bootsma 2022
- Height errors scale at 1 cm H₂O ≈ 0.74 mmHg. Ten centimetres is 7 mmHg, which is the difference between a normal and a markedly raised right atrial pressure. Bootsma 2022
- Transducer above the axis under-reads; below it over-reads. Bed height changes, head-of-bed elevation and a transducer clipped to a pole are the usual culprits.
- Re-zero before any simultaneous recording (PCWP against LVEDP, RV against LV) so a few mmHg of offset is not read as a gradient or as discordance. Nishimura 2012
- The small numbers suffer most: RAP, wedge and the transpulmonary gradient. The same 7 mmHg barely changes a PA systolic of 70.